1, connecting the appropriate nozzle;
2, press the control button to the first file;
3, perpendicular to the Pipette nozzle into the liquid under the 1-6mm (depending on the size of the pipette volume);
0.1-10ul volume Plastic Pipettes into 1-2mm under the surface
2-200ul volume Pipettes into the liquid under the 2-3mm
1-5ml volume pipette into the liquid under the 3-6mm
Note: In order to measure accurately the nozzle can be pre-washed three times, that is repeated suction of liquid 3 times.
4, the control knob slowly slide back;
5, pipette out of the liquid in front of a little wait 1-3 seconds;
1000ul the following pause one second
5-10ml pause 2-3 seconds.
6, slowly remove the nozzle, the nozzle wall to ensure no liquid. Drainage operation:
1, the nozzle at an angle against the wall of capacity;
2, slowly press the control knob to the first file and wait about 1 - 3 seconds;
3, the control knob to the second file according to the process, the nozzle will drain the remaining liquid;
4, slow controls;
5, press the eject button to eject the nozzle.